XPS/XCT Series TransducersTechnology based. Customer driven.December 1996Instruction ManualPL-52133455210Rev 1.1
DO’S AND DON’TSInstallation shall only be performed by qualified personnel and in accordance with local governing regulations.Do not route cable openl
APPLICATIONS The transducer is to be used only in the manner outlined in this instruction manual.The transducer requires no cleaning or maintenance.L
LIQUID APPLICATIONS - SUBMERGENCEIn applications where flooding is possible the transducer▼ can be fitted with a submergenceshield. The shield acts a
LIQUID APPLICATIONS - STANDPIPESIn many applications access must be made via a standpipe. In such cases, Milltronics canprovide factory bonded flanged
LIQUID APPLICATIONS - VOLUME1. Beam should not detect bin bottom. If this occurs use range extension parameters (on transceivers where available) to o
LIQUID APPLICATIONS - WATER / WASTEWATERDIFFERENTIAL LEVELPUMP CONTROLSEWAGE LIFTPL-521 15
SOLIDS APPLICATIONS - TYPICAL3. Transducer too close to material inlet. Falling material will intersect sound beam and cause erroneous rea
SOLIDS APPLICATIONS - SPECIALSTORAGE BIN WITH AGITATOR1. Transducer should be kept away from infeed.2. Where agitators are in use, use the Agitator Di
Thank you for purchasing Milltronics’ products. We endeavour to design equipment thatis simple to use and reliable in its operation, with the aim of
ABOUT THE TRANSDUCERThe Echomax XPS / XCT series of transducers operates in association with Milltronics ultrasonic level monitoring products.The tran
SPECIFICATIONS* approximate shipping weight of transducer with standard cable length▼EMC performance available upon requestKynar® is registered trade
OUTLINE AND DIMENSIONSradiating faceoptional submergence shieldrefer to associated instructionsoptional bonded flangerefer to associated instructionss
MOUNTINGDO’S AND DON’TSMount the transducer so that it is above the maximum material level by at least the blankingvalue. Refer to the associated tran
MOUNTING - LIQUID APPLICATIONSFlexible ConduitBracketSubmersiblePlywoodrigid metal conduitcouplingtransducersafetychainflexibleconduitsteel channelcou
MOUNTING - LIQUID APPLICATIONS (cont’d)nipple welded to blind flangecustomer flange, flat face onlyfactory flanged transducerboltgasketnutcouplingCust
INTERCONNECTIONDirect Connection2 Wire ExtensionCoaxial Extensionblkwhtdrain / shieldMilltronics transceiver ( typical )blkdrain / shieldwhtextend c
Comments to this Manuals